Model Number Model 5600

ELITech Model 5600 Osmometers

Specifications

  • Sample Volume: 10 mL
  • Measurement Range: Typically 20 to 3200 mmol/kg* (up to 3600 mmol/kg with extended range osmometer) @ 25 °C ambient.
  • Measurement Time: 90 seconds.
  • Resolution: 1 mmol/kg.
  • Repeatability: 2 mmol/kg Standard Deviation.
  • Linearity: 2% of reading from 100 to 2000 mmol/kg. 2% of reading from 100 to 2000 mmol/kg.
  • Calibration: Automatic using Opti-Mole osmolality standards.
  • Readout: 240 x 128 pixel backlit LCD.
  • Operating Temperature: This device has been designed for indoor use only between 15 and 37 °C
  • Storage Temperature: 0 to 60 °C.
  • Line Voltage: 100 to 240 Volts AC @ 50-60 Hz. Transient.
  • Power: 40 Watts maximum.
  • Size (H x W x D): 20 cm (8”) x 28 cm (11”) x 36 cm (14”).

Product Description

The Vapro® Vapor Pressure Osmometer is now the most intelligent osmometer ever produced. This unit brings unprecedented simplicity, convenience, and accuracy to the routine determination of somolality to help meet the demands placed on todays clinical and research laboratory.

Founded in 1967, D.A.I. Scientific has provided solutions for the following companies:

Manufacturers Trust D.A.I. Scientific

D.A.I. is proud to continue a 10+ year partnership with ELITech Group/Wescor to offer the best FreezePoint and Vapor Pressure osmometers available.

Proven performance is the cornerstone of ELITech Group products. The Vapor Pressure osmometer brings unprecedented simplicity and accuracy to the determination of osmolality in the modern research lab.

The ELITech Group enables better medical decisions by delivering high-value diagnostic solutions to laboratories operating closer to the patient. Our mission is to improve patient care by developing market-leading diagnostic products that enable physicians to more rapidly and accurately determine the course of treatment.

ELITech Group is committed to providing laboratories with superior support that customers can count on.

Product Features

Osmolality in any type of sample: fluids, viscous, and even tissue*
Sample volume of only 10 microliters
Run time of only 90 seconds
Controller operation is simple, menu driven, streamlined user interface for quick acces to menu items
User-selectable measurement modes: single sample, auto repeat mode, averaging mode for precision measurements, delayed start mode for samples requiring longer equilibration times
Superb accuracy that is unsurpassed by any other method with a error rate of less than 1%
Self cleaning thermocouple reduces maintenance and improves accuracy
Easy calibration with the push of a button, automatically sets the calibration parameters